Main features :
|
The SMP 10 is a secondary standard pyranometer that combines the sensor technology from the CMP 11, the Smart interface from the SMP 11 and the low maintenance from the CMP 10.
The SMP 10 has an internal drying cartridge that will last for at least 10 years if the housing is not opened. This minimizes maintenance significantly.
The interval for dome cleaning can be extended, and the quality of measurements maximized, by fitting SMP 10 with the CVF 4 ventilation unit.
The SMP 10 has Modbus® interface, amplified analogue output, improved response time and temperature corrected measurement data. The wide and low power supply range from 5 to 30 VDC makes integration in meteorological and solar energy stations easy. The SMP 10 is protected against over voltage, reversed polarity and short circuiting.
Thanks to the standardized output and connections of every SMP 10, exchanging instruments for recalibration is easy.
SmartExplorer Windows™ software for data logging, display of data and Modbus® address setting is provided as standard.
Specifications
| Spectral range (50% points) | 285 to 2800 nm |
| Response time (63%) | 0,7 s |
| Response time (95%) | 2 s |
| Zero offset - thermal radiation (at 200 W/m²) |
< 7 W/m² |
| Zero offset - temperature change (5 K/h) |
< 2 W/m² |
| Directional response (up to 80° with 1000 W/m² beam) | < 10 W/m² |
| Temperature dependence of sensitivity (-20ºC to +50ºC) | < 1% |
| Analogue output (-V version) | 0 - 1 V (-200...2000 W/m²) |
| Analogue output (-A version) | 4 - 20 mA (0...1600 W/m²) |
| Digital output | 2-wire RS-485 |
| Overall accuracy | ± 2% |